Estimasi Dosis Serap dengan Metode Size Specific Dose Estimation (SSDE) dan Risiko Radiasi pada Pemeriksaan CT-Scan Kepala di Rumah Sakit Daerah Mangusada Kabupaten Badung

Authors

  • Eka Saputra Prodi Pendidikan Fisika, FMIPA, Universitas Udayana
  • Gusti Ngurah Sutapa Prodi Pendidikan Fisika, FMIPA, Universitas Udayana
  • I Nyoman Paranditayana Rumah Sakit Daerah Mangusada

DOI:

https://doi.org/10.29408/kpj.v7i1.12985

Keywords:

Effective Dose, SSDE, cancer risk factors, CT-Scan head, CTDIvol, and DLP

Abstract

CT scan -Scan Those originating from X-rays on the head are very risky considering that the tissue in the brain has a greater sensitivity to ionizing radiation when compared to other organs in the human body. Problems will then arise when the use of CT-Scan sourced from X-rays can pose a radiation risk. Based on research conducted by Mathews et al. (2013) reported that at least 680,211 patients who had performed CT-Scan, 3,150 patients were diagnosed with cancer. There is a more accurate method in estimating the dose received by patients, namely using the SSDE. The dose estimation method using SSDE considers geometric factors (body size). The purpose of this study was to see how weight, gender, and age affect SSDE, using 80 patients classified into 4 age groups (children, adults, elderly, and seniors). The results showed that the average effective dose for children was 4.065 mSv for males and 3.706 mSv for females with a significant effect of weight, sex, and age on SSDE because theFCount as big4,640    FTableof 2.215

References

Miniati, K., Sutapa, G, N., dan Sudarsana, I, W, B, 2017, Uji Kelayakan Pesawat Sinar-X Terhadap Proyeksi PA (Postero-Anterior) dan LAT (Lateral) Pada Teknik Pemeriksaan Foto Thorax. Jurnal Buletin Fisika. 18 (1): 27-31.

Mubarok, S, 2015, Perbandingan CT Dose Index (CTDI) antara Panel dan 43 Pengukuran berbasis Parameter dan Kualitas Citra. Skripsi, Program Studi Fisika, Fakultas Matematika Dan Ilmu Pengetahuan Alam, Universitas Indonesia, Depok.

Mulyono, N, P, N., Hidayanto, E., Arifin, Z., dan Anam. C., 2014, Analisis Distribusi Computed Tomography Dose Index (CTDI) Pada Body Phantom. Youngster Physics Journal, 3 (4): 357-362.

Noor dan Normahayu, 2014, Dosis Radiasi Dari Tindakan CT-Scan Kepala, Journal of Environmental Engineering and Sustainable Techniology, 1 (2): 84-91.

Rima, M, K., 2020, Estimasi Dosis Radiasi dan Faktor Risiko pada Pemeriksaan Computed Tomorapy Scan Abdomen Di Rumah Sakit Islam (RSI) Sultan Agung Semarang, Skripsi, Program Studi Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Islam Negeri Walisongo,Semarang

Syaja’ah, K, N., 2019, Estimasi Dosis Efektif Radiasi CT Scan Kepala Pada Pasien Anak Menggunakan Metode Size Spesific Dose Estimation (SSDE), Tesis, Program Studi Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Institut Teknologi Bandung, Bandung.

American Association of Physicists in Medicine, 2011, Size−Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations. AAPM Report, 204, College Park, MD.

Anam, C., Fujibuchi, T., Haryanto, F., Widita, R., Arif, I., and Dougherty, G., 2019, An evaluation of computed tomography dose index measurements using a pencil ionisation chamber and small detectors, J Radiol Prot, 39 (1): 112–124

Anam, C., Haryanto, F., Widita, R., Arif, I., Fujibuchi, T., and Toyoda, T., 2017, Dougherty G. Scatter index measurement using a CT dose profiler. Journal of Medical Physics and Biophysics, 4 (1): 95-102.

Aryawijayanti, R., Susilo, dan Sutikno, 2015, Analisis Dampak Radiasi Sinar-X Pada Mencit melalui Pemetaan Dosis Radiasi Di Laboratorium Fisika Medik, Jurnal MIPA, 38 (1): 25-30.

Downloads

Published

2023-04-30

Issue

Section

Articles