Analisis Diagnostic Reference Level (DRL) Pasien Pada Pemeriksaan Abdomen Kontras dan Nonkontras Dengan Pesawat CT Scan Di RSUD Bali Mandara

Authors

  • Venesia Sesilia Br Ginting Prodi Fisika, FMIPA, Universitas Udayana
  • Gusti Ngurah Sutapa Prodi Fisika, FMIPA, Universitas Udayana
  • I Gusti Agung Ayu Ratnawati Prodi Fisika, FMIPA, Universitas Udayana
  • I Komang Widiana RSUD Bali Mandara

DOI:

https://doi.org/10.29408/kpj.v7i1.18985

Keywords:

Diagnostic Refence Level, CTDIvol, DLP, CT Scan, abdomen

Abstract

In this study, data were obtained from 35 patients aged 27-76 years. The current used varies from 46-185 mAs and a constant tube voltage of 130 kV, with different clinical symptoms. The patient's DRL value on contrast and non-contrast abdominal examination was determined by calculating the third quartile value. On non-contrast abdominal examination, the DRL CTDIvol value was 10.62 mGy and the DLP DRL was 607.76 mGy.cm and on the non-contrast abdominal examination, the DRL CTDIvol value was 32.73 mGy and the DRL DLP value was 1825.10 mGy.cm. These results indicate that the DRL CTDI vol and DRL DLP values on contrast abdominal examination at Bali Mandara Hospital were still below the actor's limits set by BAPETEN, while on abdominal examination the DRL CTDI vol and DRL DLP values exceeded the limits set by BAPETEN. This is caused by exposure which is done three times. Therefore, it is necessary to restore both the Patient Operational Standards (SOP) and the exposure factor on an abdominal contrast CT scan at Bali Mandara Hospital.

References

Bapeten, 2016, Keselamatan Radiasi Dalam Penggunaan Pesawat Sinar-X Radiologi Diagnostik Dan Intervensional. Jakarta: Bapeten.

Ega D., Eva A., Etika, Syaphira F. N., dan Eko H., 2021, Analisis Perbandingan Diagnostic Reference Level (DRL) Modalitas CT Scan Sebagai Upaya Optimasi Proteksi dan Keselamatan Radiasi di Berbagai Negara. Berkala Fisika. Vol. 24 (3): 100-108

Elshaday S. B. S., Sutapa G. N., Sudarsana W. B., 2020, Analisis Dosis Radiasi Pasien Pada Pemeriksaan CT Scan Menggunakan Aplikasi Si-INTAN. Buletin Fisika. Vol. 21 (2): 53 – 59.

Eri H., Dewi K., 2015, Dosis Pasien Pada Pemeriksaan Rutin Sinar-X Radiologi Diagnostik. Jurnal Sains dan Teknologi Nuklir Indonesia Indonesian Journal of Nuclear Science and Technology. Vol. 16, (2); 71-84

Grace S. M. S., 2018, Analisis Hasil Citra Kanker Paru Pada CT Scan Menggunakan Kontras Media, Tugas Akhir. Universitas Sumatera Utara, Medan.

Imai R., Miyazaki O., Horiuchi T., Kurosawa H., and Nosaka S., 2014, Local diagnostic reference level based on size-specificdose estimates: Assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children’s hospital. Pediatric Radiology. Vol.45;345-353.

Risalatul, L., Naily Z. J., Dezy Z.I. N., Budi P., 2019, Determination Of Local Diagnostic Reference Level (LDL) Pediatric Patients On Ct Head Examination Based On Size-Specific Dose Estimates (SSDE) Values, Journal of Vocational Health Studies. Vol. 02: 127–133

Krisna P G A Y., Sutapa G N., Yuliara I M., 2021, Analisis Dosis Efektif Pada Pemeriksaan Computer Tomography (CT) Scan Kepala RSUD Sanjiwani Gianyar. Buletin Fisika. Vol. 22 (2): 53–59

Downloads

Published

2023-04-30

Issue

Section

Articles