Analisis Laju Dosis Radiasi di Area Ruang CT_Scan di Instalasi Radiologi RSU Bali Mandara

Authors

  • Else Putri Ayu Prameswari Universitas Udayana
  • Gusti Ngurah Sutapa Universitas Udayana
  • I Gusti Agung Ayu Ratnawati Universitas Udayana
  • Ni Nyoman Ratini Universitas Udayana
  • I Nengah Artawan Universitas Udayana
  • Ni Luh Putu Trisnawati Universitas Udayana

DOI:

https://doi.org/10.29408/kpj.v8i2.25211

Keywords:

Radiation dose, CT-SCAN, NBD

Abstract

Research has been carried out on radiation dose rate analysis in the CT-SCAN area of the radiology room at RSU Mandara Bali. The data used in this research is quantitative data observed from the radiation rate from the CT-SCAN aircraft. This study aims to determine the radiation dose rate in the radiology area and its rationalization based on regulations regarding radiation NBD (dose limit values) that apply to occupational safety and health. The research data is the actual radiation dose rate determined from the dose rate and calibration factors. The results of data calculations show that the highest dose rates are respectively at T2, 30 cm, 50 cm and 100 cm (door) with values of 1.95, 0.94 and 0.05 μSv/year, while the lowest dose rates are respectively at T4, 30 cm, 50 cm and 100 cm (waiting room) with values of 0.08, 0.06 and 0.05 μSv/year. The research results show that the dose rate at each measurement point is still considered safe for radiation workers where the NBD received is still below 20 μSv/year, while the dose rate is classified as safe for the general public at measuring distances of 50 cm and 100 cm, where the NBD received is below 1 μSv/year assuming the community concerned acts as a patient or patient's family with an interest in the radiology room area.

References

Akhadi, M., 2000. Dasar-Dasar Proteksi Radiasi, Jakarta: Erlangga.

Anggarini, R., Muslim, M., & Mutanto, A. 2019. Analisis Sebaran Radiasi Hambur Di Sekitar Pesawat Sinar-X Pada Pemeriksaan Tomografi Ginjal. Jurnal Ilmiah GIGA, 17(2), 63-69.

Badan Pengawas Tenaga Nuklir. 2013. Keputusan Kepala Badan Pengawas Tenaga Nuklir Nomor 4 Bhayangkara Makassar. Fakultas Sains Dan Teknologi UIN Alauddin Makassar.

Badan Pengawas Tenaga Nuklir. 2003. Pedoman Dosis Pasien Radiodiagnostik. Keputusan Kepala Badan Pengawas Tenaga Nuklir nomor 01-P /Ka-BAPETEN/ I-03.

Bontrager, K.L., 2014. Text Book of Radiographyc and Related Anatomy Seventh Edition. St. Louis: Mosby Inc.

Departemen Kesehatan Republik Indonesia. 1999. Pedoman Peningkatan Quality Assurance. Fasilitas Pelayanan Radiologi, Jakarta.

Felayani, F., Rosidah, S., & Budiwati, T. 2016. Analisis Kebutuhan Tenaga Radiografer Berdasarkan Beban Kerja di Instalasi Radiologi RSUD Kabupaten Batang. RadX: Jurnal Ilmiah Radiologi, 1(1).

Ferusgel, A., & Berutu, A. 2018. Faktor Yang Mempengaruhi Tindakan Keselamatan Radiasi Sinar-X di Unit Radiologi Rumah Sakit Putri Hijau Medan. Journal of Borneo Holistic Health, 1(2), 264-270.

Hafid, T. 2012. Analisis Nilai Noise Citra CT Scan dengan Variasi Filter dan faktor Eksposi. Tugas Akhir. FMIPA Universitas Hasanuddin, Makassar

Hasmawati, H. 2016. Analisis Dosis Paparan Radiasi Sinar-X di Unit Radiologi RS. Bhayangkara Makassar. Doctoral dissertation, Universitas Islam Negeri Alauddin Makassar.

Ilmi, H., & Rochmayanti, D. 2018. Pengukuran Laju Paparan Radiasi dan Efektivitas Dinding serta Perisai Radiasi Ruang Panoramik. Jurnal Radiografer Indonesia. 1(1): 81-84.

Irnawati. 2018. Studi Dosis Radiasi Pada Pemeriksaan Computer Tomography (CT) Scan Dengan Nilai Computer Tomography Dose Index (CTDI) Di Rumah Sakit Bhayangkara Makassar. Fakultas Sains Dan Teknologi UIN Alauddin Makassar. Proteksi Dan Keselamatan Radiasi Dalam Pemanfaatan Tenaga Nuklir.

Malaka, T. 2017. Analisis Faktor Risiko Paparan Radiasi Sinar-X terhadap Perubahan Jumlah Limfosit pada Radiografer di Kota Palembang. Jurnal Kedokteran dan Kesehatan: Publikasi Ilmiah Fakultas Kedokteran Universitas Sriwijaya, 4(1), 1-7.

Maryanto, D., & Solichin, A. Z. 2008. Analisis Keselamatn Kerja Radiasi Pesawat Sinar-X di Unit Radiologi RSU Kota Yogyakarta. SEMNAS IV SDM Teknologi Nuklir.

Monita, R. M. R. 2021. Analisis Penerapan Keselamatan Radiasi Sinar-X Pada Pekerja Radiasi Di Instalasi Radiologi Rumah Sakit Pekanbaru Medical Center (PMC) Tahun 2020. Media Kesmas (Public Health Media), 1(1), 26-39.

Ramadhani, P., 2006. Elektronika Kedokteran CT Scanner. Jurusan Elektro, Fakultas Teknik, Universitas Hassanudin: Makassar.

Rosyida, N. 2016. Pengukuran Laju Dosis Paparan Radiasi Eksternal di Area Radioterapi RSUD Dr. Saiful Anwar Malang. Journal of Chemical Information and Modeling, 31-40.

Seeram, E., 2001. Computed Tomography Physical Principles, Clinical Applications and Quality Control. Philadelphia: W.B. Saunders Company.

Sinaga, R. A., 2018. Analisis Paparan Dosis Radiasi Jarak Aman Petugas Dan Pasien Lain di Ruang ICU.

Siregar, E. S., Sutapa, G. N., & Sudarsana, I. W. B. 2020. Analisis Dosis Radiasi Pasien Pada Pemeriksaan CT Scan Menggunakan Aplikasi Si-INTAN. Buletin Fisika, 21(2), 53-59.

Sugiantoro, S., Has, D. F. S., KM, S., & Epid, M. 2023. Analisis Faktor Risiko Keselamatan pada Instalasi Radiologi di Rumah Sakit Medika Mulia Tuban. Journal of Public Health Science Research, 3(2), 32-45.

Supriyono, P., Rahim, A. H., & Murni, T. W. 2018. Keamanan Peralatan Radiasi Pengion Dikaitkan Dengan Perlindungan Hukum Bagi Tenaga Kesehatan Di Bidang Radiologi Diagnostik. Soepra, 3(1), 102–116.

Trikasjono, T., Marjanto, D., & Timorti, B. 2009. Analisis Keselamatan Pesawat Sinar-X di Instalasi Radiologi Rumah Sakit Umum Daerah Sleman Yogyakarta. Sekolah Tinggi Teknologi Nuklir-Badan Tenaga Nuklir Nasional. Yogyakarta.

Trikasjono, T., Hanifasari, K., & Suhendro, B. 2015. Analisis Paparan Radiasi Lingkungan Ruang Radiologi di Rumah Sakit dengan Program Delphi. Jurnal Teknologi Elektro, 6(3), 142458.

Yuliamdani, R., Sahara, S., & Fuadi, N. 2020. Pengujian Paparan Radiasi Sinar-X Di Unit Radiologi RSUD Kota Makassar. JFT: Jurnal Fisika dan Terapannya, 7(1).

Downloads

Published

2024-08-04

Issue

Section

Articles