Penentuan Nilai Computed Tomography Dose Index (CTDIvol) dan Dose Length Product (DLP) Pesawat CT-Scan kepala pada kasus Stroke Non Hemoragik (SNH)

Determination of Computed Tomography Dose Index (CTDIvol) and Dose Length Product (DLP) values of head CT-Scan aircraft in cases of Non Hemorrhagic Stroke (SNH)

Authors

  • Nabilla Fatimah Udayana University
  • Gusti Ngurah Sutapa Universitas Udayana
  • Wayan Balik Sudarsana Universitas Udayana
  • Ni Nyoman Ratini Universitas Udayana
  • Putu Suardana Universitas Udayana
  • I Ketut Putra Universitas Udayana

DOI:

https://doi.org/10.29408/kpj.v8i2.25599

Keywords:

CT Scan, DRL, Stroke Non Hemoragik

Abstract

A study was conducted on the determination of Computed Tomography Dose Index (CTDIvol) and Dose Length Product (DLP) values of head CT-Scan aircraft in cases of Non Hemorrhagic Stroke (SNH). The study was conducted at the Radiology Installation of Prof.Dr. I.G.N.G. Ngoerah Denpasar Hospital using secondary data from head CT-Scan examination results consisting of age, gender, kV, mAs, slice thickness, rotation time, CTDIvol and DLP in patients aged 40-80 years. This study aims to determine the local DRL value at Prof.Dr. I.G.N.G. Ngoerah Denpasar Hospital does not exceed the national DRL value. It was found that in female patients the CTDIvol value in male patients was 54.9 mGy and in female patients it was 53.11 mGy. While the DLP value of male patients was 1143 mGy.cm and in female patients it was 1118 mGy.cm. Based on the results of the analysis with a one-way t-test, it can be seen that the DRL values of CTDIvol and DLP in both male and female patients do not exceed the standards set by BAPETEN No.1211/K/V/2021, so it can be said that the CT-Scan at the Radiology Installation of Prof.Dr. I.G.N.G. Ngoerah Denpasar Hospital is still suitable for use in examinations, especially head examinations.

References

Anam, C., Haryanto, F., Widita, R., Arif, I., & Dougherty, G. (2016). A fully automated calculation of size-specific dose estimates (SSDE) in thoracic and head CT examinations. Journal of Physics: Conference Series, 694(1). https://doi.org/10.1088/1742-6596/694/1/012030

BAPETEN. (2019). PERATURAN BADAN PENGAWAS TENAGA NUKLIR REPUBLIK INDONESIA TAHUN 2019 TENTANG KESELAMATAN RADIASI DALAM PENGGUNAAN IRADIASI DENGAN IRADIATOR.

Ega, D., Anggun, E., Faza, N., & Eko, H. (2021). Analisis Perbandingan Diagnostic Reference Level (Drl) Modalitas Ct Scan Sebagai Upaya Optimasi Proteksi Dan Keselamatan Radiasi Di Berbagai Negara. Berkala Fisika, 24(3), 100–108.

Gunawan, A. A. . (2023). Statistika Parametrik, Non Parametrik dan Multivariat.

Henderson, K. M., Clark, C. J., Lewis, T. T., Aggarwal, N. T., Beck, T., Guo, H., Lunos, S., Brearley, A., Mendes De Leon, C. F., Evans, D. A., & Everson-Rose, S. A. (2013). Psychosocial distress and stroke risk in older adults. Stroke, 44(2), 367–372. https://doi.org/10.1161/STROKEAHA.112.679159

Matsubara, K. (2017). Computed Tomography Dosimetry: From Basic to State-of-the-art Techniques. Medical Physics International, 5(1), 61–67.

Meilinda, T., Hidayanto, E., & Arifin, Z. (2014). Pengaruh Perubahan Faktor Eksposi Terhadap Nilai CT Number. Youngster Physics Journal, 3(3), 269–278.

Riskesdas. (2013). Riset Kesehatan Dasar (RISKESDAS) 2013. Laporan Nasional 2013, 1. http://www.dof.gov.my/en/c/document_library/get_file?uuid=e25cce1e-4767-4acd-afdf-67cb926cf3c5&groupId=558715

Sofiana, L., Noor, J. A. E., Normahayu, I., & Sinar-x, K. (2013). Estimasi Dosis Efektif Pada Pemeriksaan Multi Slice Ct-Scan Kepala Dan Abdomen Berdasarkan Rekomendasi Icrp 103. Fisika FMIPA Universitas Brawijaya, 1, 1–5.

Sugiyono, P. D. (2006). Statistika Untuk Penelitian. CV ALFABETA Bandung.

Susanto, W. (2018). Penentuan Nilai Diagnostic Reference Level (DRL) CT-Scan Untuk Pemeriksaan Kepala dan Dada Pasien Dewasa.

Yogantara, P. G. A. K., Sutapa, G. N., & Yuliara, I. M. (2021). Analisis Dosis Efektif Pada Pemeriksaan Computer Tomography (CT) Scan Kepala Di RSUD Sanjiwani Gianyar Effective Dose Analysis on Computer Tomography (CT) Head Scan at Gianyar Sanjiwani Hospital. Buletin Fisika, 22(2), 53–59.

Downloads

Published

2024-08-08

Issue

Section

Articles