Pengaruh Variasi Arus Tabung (mA) Terhadap Kualitas Citra Pada Pesawat Computed Tomography (CT) Scan di RSUD Bali Mandara
DOI:
https://doi.org/10.29408/kpj.v9i1.29378Keywords:
CT-Scan, Tube Current, Image Quality, SNR, CNR.Abstract
Computed Tomography (CT) Scan is one of the essential diagnostic tools in the medical field. CT scan image quality is represented by the Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR). This study was conducted to understand the effect of tube current variation on image quality in CT scan examinations. The CT scan parameters were set at an X-ray tube voltage of 120 kV, an exposure time of 1 second, and tube current variations of 100 mA, 200 mA, 300 mA, 400 mA, and 500 mA. From 25 image data obtained from scanning the Tos phantom on the CT scan, data were collected on the average signal intensity of the material, the average signal intensity of the background, and the standard deviation of the background signal intensity obtained from the Region of Interest (ROI) readings on each scanned phantom image. The Pearson correlation test results between tube current and SNR values showed that the Pearson correlation coefficient for polypropylene, nylon, acrylic, delrin, and air were 0.993, 0.992, 0.993, 0.992, and 0.992, respectively. The linear regression correlation coefficient (r) for polypropylene, nylon, acrylic, delrin, and air were also 0.993, 0.992, 0.993, 0.992, and 0.992, respectively. Similarly, the Pearson correlation test results between tube current and CNR values showed that the Pearson correlation coefficient for polypropylene, nylon, acrylic, delrin, and air were 0.992, 0.992, 0.993, 0.992, and 0.992, respectively. The linear regression correlation coefficient (r) for polypropylene, nylon, acrylic, delrin, and air were also 0.992, 0.992, 0.993, 0.992, and 0.992, respectively. These results indicate that for all phantom materials, the tube current value has a strong positive correlation with both SNR and CNR. The SNR and CNR values increase linearly with increasing tube current, suggesting that a higher tube current can improve CT scan image quality.
References
Anggarani, N. K. N., Suryatika, I. B. M., dan Sutapa, G. N. (2023). Effect of Slice Thickness Variation on Spatial Resolution of CT-Scan Device Images at the Bali Mandara Hospital Radiology Installation. Asian Journal of Medicine and Health,21(10),263–268. https://doi.org/10.9734/ajmah/2023/v21i10901.
Dewi, P. S., Ratini, N. N., dan Trisnawati, N. L. P. (2022). Effect of X-ray Tube Voltage Variation to Value of Contrast to Noise Ratio (CNR) on Computed Tomography (CT) Scan at RSUD Bali Mandara. International journal of physical sciences and engineering,6(2),82–90. https://doi.org/10.53730/ijpse.v6n2.9656.
Hutami, . A. P. A., Sutapa, G. N., and Paramarta, I. B. A. (2021). Analisis Pengaruh Slice Thickness Terhadap Kualitas Citra Pesawat CT Scan Di RSUD Bali Mandara. BuletinFisika,22(2),77–83. https://doi.org/10.24843/BF.2021.v22.i02.p04.
Kusumaningsih, L. P. R., Suryatika, I. B. M., Trisnawati, N. L. P., dan Irhas, R. (2023). Pengaruh Slice Thickness terhadap Signal to Noise Ratio (SNR) dari Hasil Penyinaran CT Scan di RSUP Prof. Dr. I.G.N.G Ngoerah. Kappa Journal, 7(2), 326– 330. https://doi.org/10.29408/kpj.v7i2.7675.
Meilinda, T., Hidayanto, E., Arifin, Z. 2014. Pengaruh Perubahan Faktor Eksposi Terhadap Nilai CT Number. Youngster Physics Journal. Vol 3, No. 3: 269-278. https://ejournal3.undip.ac.id/index.php/bfd/article/view/5945
Noveranty, A., Purwaningsih, S., & Fendriani, Y. (2024). 35155-Article Text-108444-1-10-20240805. 9(3), 53–59. https://onlinejournal.unja.ac.id/jop/article/view/35155
Papageorgiou, S. N. (2022). On Correlation Coefficients and Their Interpretation. Journal of Orthodontics,49(3),359–361. https://doi.org/10.1177/14653125221076142.
Satwika, L. G. P., Ratini, N. N., dan Maghfirotul, I. (2021). Pengaruh Variasi Tegangan Tabung Sinar-X terhadap Signal to Noise Ratio (SNR) dengan Penerapan Anode Heel Effect menggunakan Stepwedge. Buletin Fisika,22(1),20–28. https://doi.org/10.24843/BF.2021.v22.i01.p04.
Seeram, E. (2018). Computed Tomography: A Technical Review. Radiologic technology, 89(3), 279CT-302CT.
Wibowo, N. P. E dan Sunarno. 2016. Uji Profisiensi Citra Hasil Eksposi Sistem Radiografi Digital di Laboratorium Fisika Medik Unnes. Unnes Physics Journal. Vol. 1, No. 2: 527-539. http://journal.unnes.ac.id/sju/index.php/upj
Downloads
Published
Issue
Section
License
Copyright (c) 2025 Kappa Journal

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Semua tulisan pada jurnal ini menjadi tanggungjawab penuh penulis. Jurnal Kappa memberikan akses terbuka terhadap siapapun agar informasi dan temuan pada artikel tersebut bermanfaat bagi semua orang. Jurnal Kappa dapat diakses dan diunduh secara gratis, tanpa dipungut biaya, sesuai dengan lisensi creative commons yang digunakan