Analisis Electron Beam Output Constancy pada Pesawat Linear Accelelator (LINAC) Elekta Precise di Instalasi Radioterapi RSUP Sanglah Denpasar
DOI:
https://doi.org/10.29408/kpj.v6i2.6325Keywords:
Electron beam output constancy, LINAC, Deviation,Abstract
An analysis of the electron beam output constancy has been conducted on the Elekta Precise Linear Accelerator (LINAC) aircraft at the Radiotherapy Installation at Sanglah Hospital, Denpasar. The energy variations used are 8 MeV and 10 MeV. The analysis of the electron beam output constancy is determined by the standard deviation value obtained. The results of this study were obtained at an energy of 8 MeV with a large deviation value of 0.083% - 1.041%, and an energy of 10 MeV with a large deviation value of 0.022% - 0.695%. Based on the calculation results that the deviation value is still below the tolerance limit set by the AAPM. So that the LINAC Elekta Precise aircraft at the Radiotherapy Installation at Sanglah Hospital Denpasar is suitable to operate to receive patients.References
Aditya B. K., Elia, Djakaria, H. M., Amalia, T., Arianty, D., 2012. Karakteristik dan Profile Electron Beam Therapy. Jurnal of The Indonesian Radiation Oncology Society, 3 (2), Juli 2012, 55-64.
American Association of Physicists in Medicine, 1994, Comprehensive QA for Radiation Oncology : Report of AAPM Radiation Therapy Committee Task Group-40, Medical Physics, New York.
Darmawati, & Suharni. (2012). Implementasi Linear Accelerator Dalam Penanganan Kasus Kanker. Prosiding Pertemuan dan Presentasi Ilmiah Teknologi Akselerator dan Aplikasinya (hal. 36-47). Yogyakarta: Program Pasca Sarjana Fisika – UGM .
Fitriatuzzakiyyah, N., Rano K.S., Irma M.P., 2017. Terapi Kanker dengan Radiasi: Konsep Dasar Radioterapi dan Perkembangannya di Indonesia, Jurnal Farmasi Klinik.
Khan, F. M. (2014). The Physics of Radiation Therapy The 4th Edition. New York: Lippincott Williams & Wilkins.
Marten P., Jumedi, J., Dewang S., Bidayatul, A., 2015. Verifikasi percentage depth dose (PDD) dan profile dose pesawat linier accelerator (LINAC) berkas electron 6 MeV, 9 MeV, 12 MeV, dan 15 MeV menggunakan water phantom. Skripsi. Universitas Hasanudin, Makassar.
Rahayu, I.I., Nurdin, W.B., dan Samad, B.A., 2015, Analisis Dosis Output Berkas Elektron Pesawat Teleterapi Linear Accelerator (Linac) Tipe Varian HCX 6540 Menggunakan TRS 398, Skripsi, FMIPA, UNHAS, Makassar.
Setiawan, H., & Widita, R. (2016). Analisis Dosis Keluaran Berkas Foton dan Elektron Energi Tinggi Pesawat Linac Elekta Precise 5991 Berdasarkan Code of Practice IAEA TRS 398. Prosiding SKF, 80-86.
Satwika, L. G. P., Ratini, N. N., & Iffah, M. (2021). Pengaruh Variasi Tegangan Tabung Sinar-X terhadap Signal to Noise Ratio (SNR) dengan Penerapan Anode Heel Effect menggunakan Stepwedge. Buletin Fisika Vol, 22(1),20-28.
Wahyuni, A.R., 2013, Analisis Hubungan Dosis Serap dengan Jarak Sumber Radiasi ke Permukaan Medium (SSD) dan Luas Lapangan Penyinaran dari Pesawat Linear Accelerator (LINAC), Skripsi, Universitas Hasanuddin, Makassar.
Vadila, M. dan Milvita, D., 2018, Analisis Keluaran Berkas Radiasi Pesawat Terapi LINAC tipe Varian CX 6264 di Rs Unand, Jurnal Fisika Unand, Vol. 7 (2) hal. 91-96.