Penentuan Dosis Efektif Organ at Risk terhadap Penyinaran Kepala dengan Pesawat CT-Scan Merek Siemens 128 Slice di RSUP Sanglah Denpasar

Authors

  • Shatriani Wulandary Sipul Udayana University
  • Ida Bagus Made Suryatika Program Studi Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Udayana
  • I Wayan Balik Sudarsana Instalasi Radiologi, RSUP Sanglah Denpasar

DOI:

https://doi.org/10.29408/kpj.v6i2.6389

Keywords:

CT-Scan, CTDIVOL, DLP, DE, Deff

Abstract

Research has been carried out on determining the effective dose of organ at risk for head irradiation using a Siemens 128 slice CT-Scan at Sanglah Hospital, Denpasar. The data used in this study are secondary data from medical records or CT-Scan archives irradiating the head, namely CTDIVOL and DLP data for male and female patients aged 26-45 years. This study aims to determine the amount of DE and Deff received by male and female patients and to determine the feasibility of using CT-Scan in the installation. The DE value for the head section is obtained by multiplying the DLP value by the conversion factor (k). The DE analysis in the head was carried out using the standard value of the Regulation of the Head of Bapeten No. Rek: LT/STI/KN 01/P2STPFRZR1/077/2016, while Deff is obtained from the CTDIVOL value multiplied by Wt. The results of the calculation of Deff received by male patients were 0.51 ± 0.01 mGy and female patients were 0.48 ± 0.01 mGy. The results of the DE calculation for male patients were 2.38 ± 0.08 mGy and female patients were 2.28 ± 0.08 mGy. The result of the head DE calculation according to BAPETEN guidelines is 2.94 mGy. Based on the results of the DE calculation on the head, the DE value is still within the normal limits issued by the BAPETEN guidelines so that the Siemens 128 Slice CT-Scan in the installation is still suitable for use for CT-Scan examination of the head.

Author Biography

Shatriani Wulandary Sipul, Udayana University

Mahasiswa  Fisika FMIPA Universitas Udayana

References

Akhadi, M. (2000). Dasar-Dasar Proteksi Radiasi, Jakarta: Erlangga.

BAPETEN. (2019). Pedoman Teknis Penyusunan Tingkat Panduan Diagnostik atau Diagnostic Reference level (DRL) Nasional, Jakarta: Pusat Pengkajian Sistem dan Teknologi pengawasan Fasilitas Radiasi dan Zat Radioaktif Badan Pengawas Tenaga Nuklir.

Fitriyani, F., Suharyana., Muhtarom., Riyatun. (2017). Pengaruh Focus Film Distance Terhadap Kesesuaian Berkas Radiasi Pada Pesawat X-Ray Simulator di Instalasi Radioterapi RSUD Dr. Moewardi Surakarta, Universitas Sebelas Maret, Surakarta.

Holm, L. E. ICRP Managing Patient Dose in Multi-Detector Computed Tomography (MDCT): ICRP 102, Annals of the ICRP, vol. 77, 2007, pp. 1-79.

Ibrahim, A. A., Bualkar., Halmar. (2018). Estimasi Dosis Efektif Pasien Bagian Abdomen dari Hasil Pemeriksaan CT Scan Merek Siemens SOMATOM, Makasar: Prodi Fisika FMIPA UNHAS.

ICRP.( 2007). Recommendations of the International Commission on Radiological Protection Publication 103, Annals of the ICRP, Elsevier Publications, Oxford, UK.

Lestari, D. P. (2017). Kondisi Faktor Ekspose (kV, mAs) Pada Radiografi Setelah Di Lakukan Adjusment Kalibrasi Pesawat Mobile. Tugas Akhir. Universitas Sumatera Utara, Medan.

Manzil, Emidatul. (2011). Faktor Fantom Dan Estimasi Dosis Efektif DariHasil Pengukuran Computed Tomography Dose Index (CTDI). Skripsi. Program Studi Fisika, Fakultas Matematika Dan Ilmu Pengetahuan Alam, Universitas Indonesia.

Nanlohy, M. B. (2011). Faktor Koreksi Geometri dalam Pengukuran Dosis pada Phantom dengan Menggunakan Metode CTDI di Udara dan CTDI pada Phantom. Skripsi Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia.

Nariswari, N.N. (2018). Analisis Variasi Faktor Eksposi dan Ketebalan Irisan terhadap CTDI dan Kualitas Citra pada Computed Tomography (CT) Scan. Jember: Prodi Fisika Universitas Jember.

Noor, J. A. E., Normahayu, I. Dosis Radiasi dari Tindakan CT Scan Kepala, Journal of Enviromental Engineering and Sustainable Technology, Vol.1, no. 2, 2014, pp. 84-91.

Rusli, M. (2017). Uji Keselamatan Paparan Radiasi Dental Sinar-X di Radiologi Atro Muhammadiyah Makassar. Skripsi. Universitas Hasanuddin, Makassar.

Rusmanto., Intanung, S. (2019). Pedoman Teknis Penyusunan Tingkat Panduan Paparan Medik atau Diagnostic Reference Level (DRL) Nasional, Pusat Pengkajian Sistem dan Teknologi Pengawasan Fasilitas Radiasi dan Zat Radioaktif Badan Pengawas Tenaga Nuklir.

Rozanah, W., Budi, S., Arifin, Z. (2015). Perbandingan Kualitas citra CT Scan pada protokol dosis tinggi dan dosis rendah untuk pemeriksaan kepala pasien dewasa dan anak. Youngster Physics Journal. 4(1):117-126.

Silvia, H., Milvita, D., Prasetio, H., Helfi. (2013). Estimasi Nilai CTDI dan Dosis Efektif Pasien Bagian Head, Thorax dan Abdomen Hasil Pemeriksaan CT Scan Merek Philips Briliance 6, Jurnal Fisika Unand, Padang., vol.2 No.2.

Downloads

Published

2022-12-30

Issue

Section

Articles